Effect of bone-filling mesh container on bone cement injected and leakage in percutaneous kyphoplasty
Author:
Affiliation:

Department of Orthopedics, The Second Affiliated Hospital, Soochow Uni⁃versity, Suzhou 215000 , China

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    [目的] 探讨骨填充囊袋(bone-filling mesh container, BFMC) 对于经皮球囊扩张椎体后凸成形术(percutaneous ky-phoplasty, PKP) 治疗骨质疏松性椎体骨折(osteoporotic vertebral compression fracture, OVCF) 中双侧骨水泥注入量及渗漏的影响。[方法] 回顾性分析本科2022 年10 月—2023 年7 月行单椎体PKP 的100 例OVCF 患者的临床资料,根据医患沟通结果,其中50 例接受BFMC-PKP 治疗(BFMC 组),另50 例接受常规双侧PKP 治疗(PKP 组)。比较临床早期结果与影像资料。[结果] BFMC 组囊袋侧的骨水泥推注量[(4.8±0.8) ml vs (2.6±0.6), P<0.001] 显著多于PKP 侧。BFMC 侧推注的骨水泥量[(4.8±0.8) mlvs (4.1±0.9), P=0.030] 显著大于PKP 组。与术前相比,两组术后3 d VAS 评分显著降低(P<0.05),相应时间点,两组VAS 评分的差异均无统计学意义(P>0.05)。BFMC 组囊袋侧骨水泥渗漏率[例(%), 5 (10.0) vs 20 (40.0), P<0.001] 显著低于对侧,BFMC 组囊袋侧的骨水泥渗漏率[例(%), 5 (10.0) vs 14 (28.0), P=0.020] 显著低于PKP 组。骨水泥渗漏较多发生于上终板、侧方以及前方。与术前相比,两组伤椎前缘高度均显著增加(P<0.05),局部后凸角均显著减小(P<0.05)。[结论] BFMC 在PKP 治疗OVCF 时可显著降低同侧的骨水泥渗漏率,但会导致对侧骨水泥注入量减少以及渗漏率的增加。

    Abstract:

    [Objective] To investigate effect of bone-filling mesh container (BFMC) in percutaneous kyphoplasty (PKP) on bilateral bonecement injection and leakage for osteoporotic compression fracture (OVCF). [Methods] A retrospective analysis was performed on 100 pa-tients who received single-segment bilateral PKP in our department from 2022.10 to 2023.7. According to doctor-patient communication, 50of them received BFMC-PKP (the BFMC group), while other 50 received routine bilateral PKP (PKP group). Early clinical results with imag-ing data were compared. [Results] The bone cement injection volume in the BFMC side group was significantly higher than that in PKP side inthe BFMC group [(4.8±0.8) ml vs (2.6±0.6) ml, P<0.001], while the amount of bone cement injected in the BFMC side was also significantlyhigher than that in PKP group [(4.8±0.8) ml vs (4.1±0.9) ml, P=0.030]. The VAS scores significantly declined in both groups 3 days postopera-tively compared with that preoperatively (P<0.05), whereas which was not significantly different between the two groups at any time points ac-cordingly (P>0.05). Bone cement leakage rate on BFMC side was significantly lower than that in PKP side in the BFMC group [case (%), 5(10.0) vs 20 (40.0), P<0.001], and which on BFMC side in the BFMC group was significantly lower than that in PKP group [case (%), 5 (10.0) vs14 (28.0), P=0.020] . Most of the bone cement leakage occurred in the upper endplate, lateral and anterior sides. The anterior vertebral heightof injured vertebrae increased significantly (P<0.05), while the local kyphotic angle decreased significantly in both groups postoperatively com-pared with those preoperatively (P<0.05). [Conclusion] In PKP for OVCF, BFMC can significantly reduce the bone cement leakage on the ipsi-lateralside,butleadtoadecreaseintheamountofbonecementinjectionandanincreaseintheleakagerateonthecontralateralside.

    参考文献
    相似文献
    引证文献
引用本文

宋方龙,单冰晨,周震涛,等. 骨填充囊袋对后凸成形骨水泥注入量及渗漏的影响[J]. 中国矫形外科杂志, 2024, 32 (21): 2002-2006. DOI:10.20184/j. cnki. Issn1005-8478.100786.
SONG Fang-long, SHAN Bing-chen, ZHOU Zhen-tao, et al. Effect of bone-filling mesh container on bone cement injected and leakage in percutaneous kyphoplasty[J]. Orthopedic Journal of China , 2024, 32 (21): 2002-2006. DOI:10.20184/j. cnki. Issn1005-8478.100786.